Wednesday, January 17, 2018
Holy crescents, Batman!
Some participants have obvious bright crescent-shaped artifacts in their standard deviation images (the examples above are from two people; both calculated from non-censored frames, after completing the HCP Minimal Preprocessing pipeline). Looking over people and runs (some participants have completed 6 imaging sessions, over months), people have the crescents or not - their presence doesn't vary much with session (scanning day), task, or movement level (apparent or real).
They do, however, vary with encoding direction: appearing in PA phase encoding runs only. Plus, they seem to vary with subject head size, more likely in small-headed people (large-headed people seem more likely to have "ripples", but that's an artifact for another day).
All that (and thanks to conversations with practiCal fMRI and @DataLoreNeuro) gave a hint: these crescents appear to be N/2 ghost artifacts.
Playing with the contrast and looking outside the brain has convinced me that the crescents do align with the edges of ghost artifacts, which I tried to show above. These are from a raw image (the HCP Minimal Preprocessing pipelines mask the brain), so it's hard to see; I can share example NIfTIs if anyone is interested.
So, why do we have the bright ghosts, what should we do about it, and what does that mean for analysis of images we've already collected? Suggestions are welcome! For analysis of existing images, I suspect that these will hurt our signal quality a little: we want the task runs to be comparable, but they're not in people with the crescent: voxels within the crescent areas have quite different tSNR in the PA and AP runs.
Holy crescents, Batman! (We've been watching the 1966 Batman TV series.)
Sunday, April 15, 2018
more crescents: with sequence variations
We have begun a set of pilot scans to see if some parameter combinations produce better subcortical and frontal signal in a reward task. So far all scans have been on a Siemens Prisma,
Of the two pilot people (so far), one has the crescent artifact and the other does not, and the appearance of the crescents in the different acquisitions is interesting. All of these images are voxelwise standard deviation, calculated over the entire run (no censoring, but extremely low motion), and on raw images (preprocessing is in progress).
First, here are sagittal views of a flat (left, scan 15) and tilted (right, scan 37) 2.4 mm isotropic run. The crescent artifact is visible in both; I marked the approximate ends with green arrows (click to enlarge). The multiband slice boundary is visible in both a fourth of the way up the image (red arrows).
Here are axial slices of the set of runs we have so far for this person. All are with the same color scaling (0 to 200); brighter is higher standard deviation. These are raw images, so the slice appearance varies quite a bit between the "flat" and "tilted" runs.
The "crescents" are visible in all PA runs, though perhaps easiest to spot with the 2p4 (2.4 mm isotropic) voxels. The slices in which the crescents appear varies between the tilted and flat acquisitions (e.g., k=31-46 for run 15_2p4flat_PA; 22-31 for run 37_2p4tilt_PA). It will be easier to compare the crescent locations after preprocessing.
The 3p0 (3.0 mm isotropic voxels) images are generally more uniform and dark than the 2p4 runs, likely reflecting improved signal-to-noise, particularly in the middle of the brain. While the large vessels are brightest in all runs (as they should be), the runs with 2.4 mm voxels (2p4) generally have a "starburst" type effect (brighter in the center, darker towards the edges), which is worrying, particularly since we want good signal in reward areas.
I will share other observations on this blog as the piloting and analyses progress. Please contact me if you'd like to run your own analyses; we'd be happy to share and are very interested in others' thoughts.
UPDATE 18 April 2018: I've wondered before if head size was a factor in which people have the crescent artifact, using the total intracranial volume measurement produced by freesurfer as a proxy. I don't have those measurements yet, but they kindly allowed me to measure their heads as if fitting them for hats, and they were nearly identical: about 58 cm for the person without the crescent artifact, and about 57 for the person shown in this person (with the artifact). Both people have a normal healthy body size; the person without the artifact was a bit shorter (around 5'2") than the other (around 5'7"). So, at least for these two pilots, external head size doesn't seem to matter for the artifact.
more acquisition parameters below the jump
Friday, December 6, 2024
"crescent" artifacts revisited
link to cortex volume?
only with DMCC or its MB4 acquisition?
but does it matter?
thoughts?
Wednesday, June 18, 2025
OHBM 2025 (let's chat about crescents!)
Friday, June 18, 2021
DMCC55B supplemental as tutorial: basic fMRI QC: temporal mean, SD, and tSNR
This is the fourth post in a series describing the DMCC55B supplemental files. The first introduces the dataset, the second questionnaire data, and the third motion. Here I'll highlight another basic fMRI QC practice: creating and checking temporal mean, standard deviation, and tSNR images. For background on these types of QC images I suggest starting with this post, especially the multiple links in the first sentence.
For DMCC55B, we preprocessed the images as both volume and surface, so the QC was also both volume and surface. I generally find QC easier to judge with unmasked volumes, but if surface analysis is planned, the surface representations must be checked as well.
During DMCC data collection we create a QC summary file for each individual participant (example here), with volumes, surfaces, and motion (realignment parameters) for every run. A first QC check for (unmasked) volume images is straightforward: do they look like brains? This ("got brains?") check doesn't apply to surfaces, since any valid gifti image will be brain-shaped. Instead, our first QC pass for the surface images is to check the temporal mean for "tiger stripes" around the central sulcus. Below is the vertex-wise means for Bas Stern AP from the example, with the blue lines pointing to the stripes. If these stripes are not visible, something likely went wrong with the preprocessing.
For ease of group comparison, the DMCC55B QC supplemental files are split by type rather than participant: QC_SD_surface.pdf for vertex-wise temporal standard deviation, QC_tSNR_surface.pdf for vertex-wise tSNR; QC_SD_volume.pdf and QC_tSNR_volume.pdf for the corresponding voxel-wise images. With eight runs for each of 55 participants these QC summary files are still rather long, but the consistent scaling (same intensity used for all participants when plotting) makes it possible to scroll through and spot trends and oddities. Such group review can be very beneficial, especially before embarking on statistical analyses.
For example, the SD images for some participants/runs are much brighter and fuzzier than others (compare f2157me and f2499cq on page 8). Broadly, the more the temporal SD images resemble maps of brain vessel density, the better the quality, though of course the resolution will vary with fMRI acquisition parameters (e.g., in the 2.4 mm isotropic DMCC images the Circle of Willis should be clearly visible). Artifacts can also be spotted, such as the ghostly "crescents" in some participant's PA images (e.g., f5001ob).
Implementation notes:
The .rnw file with each (knitr) pdf has the R code to create the temporal statistic images, as well as produce the summary .pdfs. The startup code chunk of each document has the code to make the images (from the 4d niftis and giftis produced by fmriprep), while the later chunk plots the images.
The volumetric code uses the AFNI 3dTstat and 3dcalc functions (called from R using system2()) to make the temporal mean, standard deviation, and tSNR images, which are written out as NIfTI images. Saving the statistical images is useful for compilation speed, but more importantly, so that they can be examined in 3d if closer checking is needed. I strongly suggest using existing functions from a well-established program (such as AFNI) whenever possible, for clarity and to avoid introducing bugs.
The volumetric images are plotted with base R image() rather than with my volumetric plotting function, because I wanted to plot one slice through each plane. Thus, the _volume 55B supplementals can also serve as an example of simple NIfTI image plotting. (This is my general volume-plotting knitr tutorial.)
Finally, there is a comment in the volume .rnw files that the input images are already in LPI orientation. I strongly suggest transforming NIfTI images to LPI immediately after preprocessing (e.g., with 3dresample), if they are not already, as this often avoids left/right flipping.
Unlike volumes, the surface QC code calculates the vertex-wise mean and standard deviation in R, and my gifti plotting functions. Many AFNI functions can work with giftis, but these calculations (mean, standard deviation) are so simple, and plotting surfaces from a vector so straightforward, that I decided to save the images as text rather than gifti.