Wednesday, June 18, 2025

OHBM 2025 (let's chat about crescents!)

I'll be at OHBM 2025 next week (want to meet up? drop me a line), presenting a poster about the crescent artifacts (""Crescent" artifacts with multiband fMRI acquisitions: appearance, causes, and consequences"). The poster pdf is available via OHBM's system, or at https://osf.io/pzj39.

My blog post last December covers some of the same information, but preparing the poster refined my thoughts a bit. I still don't have as clear a sense as I'd like of exactly how much the artifact affects fMRI signal quality, but am confident that there is enough likelihood of a substantial negative impact that they shouldn't be ignored. 

It seems to be a given in the MR physics literature that Nyquist ghosts appreciably degrade EPI signal to noise. Quantifying the impact in GLM results is difficult, however, especially after preprocessing, smoothing, in group analyses (artifact location and intensity varies across people), and when runs of different encoding directions are analyzed together. Qualitatively, I have seen crescents in single-subject statistical images (average activation, GLMs), but that is of course not the typical analysis. It could be informative to run a few comparison tests; for example, is the effect strength in frontal parcels (i.e., those in the path of the crescent artifact) worse in a group of participants with the artifact than without? Or in PA than AP encoding runs?

While the degree of impact is uncertain, I think there's enough evidence to recommend that crescent artifacts be one of the criteria for choosing acquisition protocols: select fMRI acquisition parameters so that crescent artifacts are minimized and/or appear in brain areas of low theoretical interest. Since the crescent artifact likely reduces BOLD signal quality somewhat, and is more often prominent in people with smaller brains, there's a risk of bias if an experimentally-important participant characteristic (e.g., age, gender) is associated with differences in head size; extra care should be taken in these cases.

Friday, December 6, 2024

"crescent" artifacts revisited

Back in 2018 I posted twice about "crescent" artifacts (first, second), and have kept a casual eye out for them since, wondering how much they might affect fMRI analysis results. 

The artifact isn't totally consistent across people, but when present, is very stable over time (i.e., sessions several years apart), and bright in temporal standard deviation QC images. The artifact's location varies with encoding direction: front for PA encoding, rear with AP encoding; the PA encoding versions tend to be much brighter and obvious than the AP. 

Below is an example of the artifact, pointed out by the pink arrows. This is the temporal standard deviation image for sub-f8570ui from the DMCC55B dataset (doi:10.18112/openneuro.ds003465.v1.0.6), the first (left, AP encoding) and second (right, PA encoding) runs of Sternberg baseline (ses-wave1bas_task-Stern), after fmriprep preprocessing:

These two runs were collected sequentially within the same session, but the artifact is only visible in the PA encoding run (right). (Briefly, DMCC used a 3T Siemens Prisma, 32-channel headcoil, CMRR MB4, no in-plane acceleration, 2.4 mm iso, 1.2 s TR, alternating AP and PA encoding runs; details at openneuro, dataset description paper, and OSF sites, plus DMCC-tagged posts on this blog.)

In the previous "crescent" posts we speculated that these could be N/2 ghosts or related to incomplete fat suppression; I am now leaning away from the Nyquist ghost idea, because the crescents don't appear to line up with the most visible ghosts. (Some ghosts are a bit visible in the above image; playing with the contrast and looking in other slices makes the usual multiband-related sets of ghosts obvious, but none clearly intersect with the artifact.) It also seems odd that ghosts would be so much brighter and change their location with AP vs. PA encoding; I am no physicist, though!

link to cortex volume?

This week I gave three lab members a set of temporal standard deviation images (similar to the pair above) for 115 participants from the DMCC dataset. The participant images were in random order, and I asked the lab members to rate how confident they were that each showed the "crescent" artifact or not. My raters agreed that 34 participants showed the artifact, and 39 did not. (Ratings were mixed or less confident on the others; I asked them to decide quickly from single pairs of QC images, not investigate closely.)

We didn't measure external head size in the participants, but did run freesurfer during preprocessing, so I used its CortexVol and eTIV statistics as proxies (a different stat better?): and the group my colleagues rated as having the artifact tended to have smaller brains than those without:

If the appearance of this artifact is indeed somewhat related to head size, then it's logical that it would (as I've observed) generally be stable over time. DMCC's population was younger adults; it'd be interesting to see if there's a relationship with a wider range of head sizes.

only with DMCC or its MB4 acquisition?

Is the artifact restricted to this particular acquisition or study? No, not somehow related to DMCC; I've checked a few DMCC participants with the artifact who later participated in other studies and they have it (or not) in all of the datasets.

To see if it's restricted to the MB4 acquisition, I looked at a few images from a different study, which also has adult participants, a 3T Prisma, 32-channel headcoil, 2.4 mm iso voxels, and CMRR sequences, but with MB6 TR 0.8 s, and PA encoding for all runs. Below are standard deviation images for three different people from this MB6 study, one run each of the same task, after fmriprep preprocessing. (I chose these three because of the artifacts; not all are so prominent.)

Since this study has all PA runs I can't directly compare artifacts across the encoding directions, but there are clearly some "crescents", and more sets of rings than typical with MB4 (which makes sense for MB6). The rings are especially obvious in person #3; some of these appear to be full-brain ghosts. I suspect the artifacts would be much clearer in subject space; I haven't looked (I'm not directly involved in the study). But a substantial minority of these participants' standard deviation images resemble #1, whose artifacts strike me as quite similar to the "crescents" in some DMCC MB4 PA images.

but does it matter?

Not all artifacts that look strange in QC images actually change task-related BOLD statistics enough to be a serious concern. (Of course, how much is too much totally depends on the particular case!) I suspect that this artifact does matter for our analyses, though, both because of where it falls in the brain and because it affects BOLD enough to be visible by eye in some cases.

The artifact's most prominent frontal location with PA runs puts it uncomfortably close to regions of interest for most of my colleagues, and is one reason I have advised shifting to all AP encoding for new studies I'm involved with. Preprocessing, motion, transformation to standard space, and spatial smoothing blurs the artifact across a wider area, hopefully diluting its effect. But the artifact's location is somewhat consistent across participants, and present in a sizable enough minority (a third, perhaps, in the datasets I've looked at), that it seems possible it could reduce signal quality in our target ROIs.

For showing that it does indeed actually affect task-related BOLD enough to matter, so far I mostly just have qualitative impressions. For example, below left is the standard deviation of one DMCC person's PA Sternberg run, with the cursors on the artifact. The right side is from averaging together (after voxel-wise normalizing and detrending) frames after pressing a button with the right hand. Squinting, the statistical image is brighter in sensible motor-related grey matter areas, marked with green. But the "crescent" may also be faintly visible, as pointed out in pink.

I can imagine quantitative tests, such as comparing the single-run (so separating AP and PA encoding runs) GLM output images from the group of participants with and without the artifact. Differences in estimates in parcels/searchlights/areas overlapping the artifact would be suggestive, particularly as the estimates vary with encoding direction and participant subgroup (with-artifact or without).

thoughts?

I'm curious, have you seen similar? Do you think this artifact is from N/2 ghosting, incomplete fat suppression, or something else? (What should I call it? "Crescent" is visually descriptive, but not standard. 😅) Seem reasonable it could be related to head size? And that it can significantly affect BOLD? Other reactions? Thanks! (And we can chat about this at my OHBM 2025 poster, which will be on this topic.)

Friday, June 18, 2021

DMCC55B supplemental as tutorial: basic fMRI QC: temporal mean, SD, and tSNR

This is the fourth post in a series describing the DMCC55B supplemental files. The first introduces the dataset, the second questionnaire data, and the third motion. Here I'll highlight another basic fMRI QC practice: creating and checking temporal mean, standard deviation, and tSNR images. For background on these types of QC images I suggest starting with this post, especially the multiple links in the first sentence.

For DMCC55B, we preprocessed the images as both volume and surface, so the QC was also both volume and surface. I generally find QC easier to judge with unmasked volumes, but if surface analysis is planned, the surface representations must be checked as well.

During DMCC data collection we create a QC summary file for each individual participant (example here), with volumes, surfaces, and motion (realignment parameters) for every run. A first QC check for (unmasked) volume images is straightforward: do they look like brains? This ("got brains?") check doesn't apply to surfaces, since any valid gifti image will be brain-shaped. Instead, our first QC pass for the surface images is to check the temporal mean for "tiger stripes" around the central sulcus. Below is the vertex-wise means for Bas Stern AP from the example, with the blue lines pointing to the stripes. If these stripes are not visible, something likely went wrong with the preprocessing.


For ease of group comparison, the DMCC55B QC supplemental files are split by type rather than participant: QC_SD_surface.pdf for vertex-wise temporal standard deviation, QC_tSNR_surface.pdf for vertex-wise tSNR; QC_SD_volume.pdf and QC_tSNR_volume.pdf for the corresponding voxel-wise images. With eight runs for each of 55 participants these QC summary files are still rather long, but the consistent scaling (same intensity used for all participants when plotting) makes it possible to scroll through and spot trends and oddities. Such group review can be very beneficial, especially before embarking on statistical analyses.

For example, the SD images for some participants/runs are much brighter and fuzzier than others (compare f2157me and f2499cq on page 8). Broadly, the more the temporal SD images resemble maps of brain vessel density, the better the quality, though of course the resolution will vary with fMRI acquisition parameters (e.g., in the 2.4 mm isotropic DMCC images the Circle of Willis should be clearly visible). Artifacts can also be spotted, such as the ghostly "crescents" in some participant's PA images (e.g., f5001ob).

Implementation notes:

The .rnw file with each (knitr) pdf has the R code to create the temporal statistic images, as well as produce the summary .pdfs. The startup code chunk of each document has the code to make the images (from the 4d niftis and giftis produced by fmriprep), while the later chunk plots the images. 


The volumetric code uses the AFNI 3dTstat and 3dcalc functions (called from R using system2()) to make the temporal mean, standard deviation, and tSNR images, which are written out as NIfTI images. Saving the statistical images is useful for compilation speed, but more importantly, so that they can be examined in 3d if closer checking is needed. I strongly suggest using existing functions from a well-established program (such as AFNI) whenever possible, for clarity and to avoid introducing bugs.

The volumetric images are plotted with base R image() rather than with my volumetric plotting function, because I wanted to plot one slice through each plane. Thus, the _volume 55B supplementals can also serve as an example of simple NIfTI image plotting. (This is my general volume-plotting knitr tutorial.) 

Finally, there is a comment in the volume .rnw files that the input images are already in LPI orientation. I strongly suggest transforming NIfTI images to LPI immediately after preprocessing (e.g., with 3dresample), if they are not already, as this often avoids left/right flipping. 



Unlike volumes, the surface QC code calculates the vertex-wise mean and standard deviation in R, and my gifti plotting functions. Many AFNI functions can work with giftis, but these calculations (mean, standard deviation) are so simple, and plotting surfaces from a vector so straightforward, that I decided to save the images as text rather than gifti.

Sunday, April 15, 2018

more crescents: with sequence variations

Back in January I posted about the "crescent" artifacts that show up in the PA runs of some people. (In a few people reversed crescents appear in the AP runs as well.) The working hypothesis is still that these are N/2 ghosts and perhaps related to insufficient fat suppression. I am still extremely interested in any advice people have for avoiding these, especially as I now have evidence that task signal is noticeably reduced in the "crescent" areas (details soon, hopefully).

We have begun a set of pilot scans to see if some parameter combinations produce better subcortical and frontal signal in a reward task. So far all scans have been on a Siemens Prisma, 64 32 channel head coil, CMRR MB4 sequences; we'll be doing the tests on a Siemens Vida as well in a few weeks. So far, scans are acquired with 2.4 or 3.0 mm isotropic voxels, either "flat" (AC-PC aligned as usual) or "tilted" (30 degrees off AC-PC); more acquisition details below the jump.

Of the two pilot people (so far), one has the crescent artifact and the other does not, and the appearance of the crescents in the different acquisitions is interesting. All of these images are voxelwise standard deviation, calculated over the entire run (no censoring, but extremely low motion), and on raw images (preprocessing is in progress).

First, here are sagittal views of a flat (left, scan 15) and tilted (right, scan 37) 2.4 mm isotropic run. The crescent artifact is visible in both; I marked the approximate ends with green arrows (click to enlarge). The multiband slice boundary is visible in both a fourth of the way up the image (red arrows).

Here are axial slices of the set of runs we have so far for this person. All are with the same color scaling (0 to 200); brighter is higher standard deviation. These are raw images, so the slice appearance varies quite a bit between the "flat" and "tilted" runs.
The "crescents" are visible in all PA runs, though perhaps easiest to spot with the 2p4 (2.4 mm isotropic) voxels. The slices in which the crescents appear varies between the tilted and flat acquisitions (e.g., k=31-46 for run 15_2p4flat_PA; 22-31 for run 37_2p4tilt_PA). It will be easier to compare the crescent locations after preprocessing.

The 3p0 (3.0 mm isotropic voxels) images are generally more uniform and dark than the 2p4 runs, likely reflecting improved signal-to-noise, particularly in the middle of the brain. While the large vessels are brightest in all runs (as they should be), the runs with 2.4 mm voxels (2p4) generally have a "starburst" type effect (brighter in the center, darker towards the edges), which is worrying, particularly since we want good signal in reward areas.

I will share other observations on this blog as the piloting and analyses progress. Please contact me if you'd like to run your own analyses; we'd be happy to share and are very interested in others' thoughts.

UPDATE 18 April 2018: I've wondered before if head size was a factor in which people have the crescent artifact, using the total intracranial volume measurement produced by freesurfer as a proxy. I don't have those measurements yet, but they kindly allowed me to measure their heads as if fitting them for hats, and they were nearly identical: about 58 cm for the person without the crescent artifact, and about 57 for the person shown in this person (with the artifact). Both people have a normal healthy body size; the person without the artifact was a bit shorter (around 5'2") than the other (around 5'7"). So, at least for these two pilots, external head size doesn't seem to matter for the artifact.

more acquisition parameters below the jump

Wednesday, January 17, 2018

Holy crescents, Batman!

Quite a few of the posts over the last year or so have arisen from things that catch my eye as I review the SMS/MB4 images we're collecting in our ongoing project, and this is another. For quick comparison, I make (with knitr; we may give mriqc a try) files showing slices from mean, standard deviation, and tSNR images for participants, runs, and sessions.


Some participants have obvious bright crescent-shaped artifacts in their standard deviation images (the examples above are from two people; both calculated from non-censored frames, after completing the HCP Minimal Preprocessing pipeline). Looking over people and runs (some participants have completed 6 imaging sessions, over months), people have the crescents or not - their presence doesn't vary much with session (scanning day), task, or movement level (apparent or real).

They do, however, vary with encoding direction: appearing in PA phase encoding runs only. Plus, they seem to vary with subject head size, more likely in small-headed people (large-headed people seem more likely to have "ripples", but that's an artifact for another day).

All that (and thanks to conversations with practiCal fMRI and @DataLoreNeuro) gave a hint: these crescents appear to be N/2 ghost artifacts.

Playing with the contrast and looking outside the brain has convinced me that the crescents do align with the edges of ghost artifacts, which I tried to show above. These are from a raw image (the HCP Minimal Preprocessing pipelines mask the brain), so it's hard to see; I can share example NIfTIs if anyone is interested.

So, why do we have the bright ghosts, what should we do about it, and what does that mean for analysis of images we've already collected? Suggestions are welcome! For analysis of existing images, I suspect that these will hurt our signal quality a little: we want the task runs to be comparable, but they're not in people with the crescent: voxels within the crescent areas have quite different tSNR in the PA and AP runs.

Holy crescents, Batman! (We've been watching the 1966 Batman TV series.)